Technical Guides
Jun 16, 2026 . 0 Comments

Pressure Transmitter Calibration and Troubleshooting Guide

author-img

industryinchina

Practical guide for pressure transmitter calibration, loop testing, diagnostic procedures, and solutions for common problems.

Pressure Transmitter Calibration and Troubleshooting Guide

Regular calibration ensures pressure transmitters provide accurate measurements for process control and safety applications.

I. Calibration Procedure

  1. Isolate the transmitter from the process using the manifold valves
  2. Connect a pressure calibrator or deadweight tester to the high pressure side
  3. Vent the low pressure side to atmosphere (for gauge pressure calibration)
  4. Apply pressure at calibration points from zero to one hundred percent in twenty-five percent steps
  5. Record the output at each point for both increasing and decreasing pressure
  6. Calculate errors and adjust trim if the error exceeds specifications

II. Troubleshooting

Output Stuck at Zero or Minimum: Check for reversed impulse lines. Verify that the transmitter is not in simulation mode.

Output Above Range: Check for overpressure condition. Verify the range configuration matches the actual process pressure. Check for sensor damage.

Drifting Output: Check for temperature changes beyond compensation limits. Verify power supply stability. Check for moisture in the electronics housing.

Noisy Output: Check for process pulsations. Increase damping. Verify that impulse lines are not partially plugged.

Leave a Comment